MDA/Malondialdehyde ELISA Kit
The MDA ELISA Kit can assay for MDA / Malondialdehyde in the following samples: serum, blood, plasma, cell culture supernatant and other related supernatants and tissues.
How our MDA / Malondialdehyde ELISA Kits Work?
The ELISA Genie ELISA (enzyme-linked immunosorbent assays) assay kits are designed for the quantitative measurement of analytes in a wide variety of samples. As today’s scientists demand high quality consistent data for high impact journals, We have developed a range of sensitive, fast and reliable ELISA kit assays to meet and exceed those demands.
This ELISA kit uses Competitive-ELISA as the method. The microtiter plate provided in this kit has been pre-coated with MDA / Malondialdehyde . During the reaction, MDA / Malondialdehyde in the sample or standard competes with a fixed amount of MDA / Malondialdehyde on the solid phase supporter for sites on the Biotinylated Detection Antibody specific to MDA / Malondialdehyde . Excess conjugate and unbound sample or standard are washed from the plate, and HRP-Streptavidin(SABC) is added to each microplate well and incubated. Then a TMB substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of a sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450 nm. The concentration of MDA / Malondialdehyde in the samples is then determined by comparing the O.D. of the samples to the standard curve.
- Product name:
MDA (Malondialdehyde) ELISA Kit
- Catalogue No.:
EU2577
- Alias:
MDA(Malondialdehyde)
- Detection method:
Competitive ELISA Coated with Antigen
- Application:
This immunoassay kit allows for the in vitro quantitative determination of MDA concentrations in serum plasma and other biological fluids.
- Size:
96T
- Range:
31.25-2000ng/ml
- Sensitivity:
< 18.75ng/ml
- Storage:
4 „ƒ for 6 months
- Recovery:
Matrices listed below were spiked with certain level of MDA and the recovery rates were calculated by comparing the measured value to the expected amount of MDA in samples.
Matrix Recovery range(%) Average(%) serum(n=5) 88-103 93 EDTA plasma(n=5) 94-104 98 UFH plasma(n=5) 87-102 91 - Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of MDA and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample 1:2 1:4 1:8 1:16 serum(n=5) 86-98% 87-105% 87-101% 87-95% EDTA plasma(n=5) 83-99% 82-101% 85-99% 82-99% UFH plasma(n=5) 93-100% 80-100% 84-97% 82-97% - CV(%):
Intra-Assay: CV<8%
Inter-Assay: CV<10% - Note: