1. The Nano-300 isa perfect full-spectrum, UV-Vis Micro-Spectrophotometer based on Nano-100 with full range of wavelength (200-800nm).
2. The Nano-300 added a new function of bacterium cell concentration test (OD600).
3. The Nano-300 uses 7 inch touch screen and is integrated with Android system, which do not need connect to a computer.
(200-800nm). It added a new function of bacterium cell concentration test (OD600). The Nano-300 only
requires 2ul sample and can test the nucleic acid, Protein quickly like the nano-100, it also equips with a
cuvette and can test the concentration of bacterium solution to know the growing level of the bacterium.
Nano-300 uses 7 inch touch screen and is integrated with Android system, which do not need connect
to a computer. And it is an ideal product for the biology laboratory which makes the life science research
Applications:
Easy Handling:
Code Description
AS-11020-00 Nano-300 DC24V
1. Easy use: Android system, 7 inch touch screen, APP soft, simple interface, easy to use
2. Micro-Volumes: Same as Nano-100, only require 0.5~2ul sample to accurate determinate
3. Long life’s Xenon flash lamp: Xenon flash lamp, 10 flashes, up to 10years, no cells or cuvetes
4. Flexible data output: The data can be printed with built-in printer, and can output via USB memory
5. OD600: OD600nm Measuring, It can measure the culture fluid concentration of bacteria etc.
6. No Expensive consumables: Direct measurement without wasteful dilitions and expensive comsumables
7. Easy & Fast: Direct, easy measurements within 8sec. No dilutions for samples and no warm up.
8. No computer required: Touch screen and Android system operation without computer
Model | Nano-300 |
Wavelength Range | 200~800nm |
Minimum Sample Size | 0.5~2.0ul |
Path Length | 0.2mm (For high concentration measwrement) 1.0mm (For ordinary) |
Light Source | Xenon flash lamp |
Detector Type | 3864-element linear silicon CCD array |
Wavelength Accuracy | 1nm |
Spectral Resolution | ?3nm (FWHM at Hg 546nm) |
Absorbance Precision | 0.003Abs |
Absorbance Accuracy | 1% (7.332Abs at 260nm) |
Absorbance Range | 0.02 – 90A (10mm equivalent) |
Detects Nucleic Acid up to | 3~4500ng/ul (dsDNA) |
Measurement Time | <10s |
Dimensions (W x D x H)mm | 210 x 268 x 181 |
Weight | 3.0kg |
Sample Pedestal Material | Aluminum alloy and Quartz fiber |
Operating Voltage | 24VDC |
Operating Power Consumption | 40W |
Standby Power Consumption | 5W |
Software Compatibility | Android System |
OD600nm Measurement: | |
Light Source | LED |
Wavelength Range | 600±8nm |
Abosorbance Range | 0~3A |
Reviews
There are no reviews yet.